ABSTRACT
Introduction: Liposuction is a consecrated procedure to improve body contouring, and various types of laser can be applied to cause lipolysis and skin retraction. The current parameters (power and energy) in the devices allow only an indirect inference of skin temperature change.
Methods: Eighty-three patients, between 17 and 75 years old, were submitted to laser lipolysis with 980nm diode, using a Flir® T540 thermocamera for temperature monitoring. Lipolysis was obtained by applying the laser to the deep subcutaneous layer and skin retraction in the deep dermis, with a velocity of 5 centimeters per second. The laser was used in the deep dermis of the abdominal flaps of patients undergoing dermolipectomy to determine the temperature that causes skin burns, totaling 27 cases.
Results: A power of 15W was used on the medial side of the arm and thigh, and 4000J average energy was applied per region. In the dorsal area and abdomen, 20W was adopted, and 6000J were applied in each quadrant. On average, the studied tissue's initial temperature was 31°C, and at the end of the laser application was 37°C. The temperature that caused the burn was 45°C.
Conclusion: The use of the thermocamera allowed a more homogeneous distribution of the laser and provided the surgeon with effective monitoring so that an adequate dose of energy is applied to the subcutaneous and deep dermis, obtaining the desired effect.
Keywords: Lipectomy; Laser; Thermography; Subcutaneous fat; Skin.
RESUMO
Introdução: A lipoaspiração é um procedimento consagrado para melhora do contorno corporal, sendo que existem diferentes tipos de laser que podem ser aplicados com o objetivo de provocar lipólise e retração da pele. Os parâmetros existentes atualmente (potência e energia) nos aparelhos permitem apenas uma inferência indireta da mudança de temperatura da pele.
Métodos: Oitenta e três pacientes, entre 17 e 75 anos, foram submetidos à laserlipólise com diodo 980nm, utilizando uma termocâmera Flir® T540 para monitorização da temperatura. A lipólise foi obtida pela aplicação do laser na camada subcutânea profunda e a retração da pele na derme profunda, com velocidade de 5 centímetros por segundo. Para determinar a temperatura que provoca queimadura na pele, o laser foi aplicado na derme profunda dos retalhos abdominais das pacientes submetidas à dermolipectomia, totalizando 27 casos.
Resultados: Na face medial do braço e da coxa foi utilizada uma potência de 15W e foram aplicados em média 4000J de energia por região. Na região dorsal e abdome foi adotada potência de 20W e foram aplicados 6000J em cada quadrante. No submento foram utilizados 10W e aplicados 1500J. Em média, a temperatura inicial do tecido estudado foi de 31°C, e ao final da aplicação do laser foi de 37°C. A temperatura que provocou queimadura foi 45°C.
Conclusão: A utilização da termocâmera permitiu uma distribuição mais homogênea do laser e propiciou ao cirurgião uma monitorização eficaz, para que seja aplicada no subcutâneo e na derme profunda uma dose adequada de energia, obtendo-se o efeito desejado.
Palavras-chave: Lipectomia; Laser; Termografia; Gordura subcutânea; Pele
ABSTRACT
Twenty-six patients, 22 females and 4 males were selected in order to evaluate the histological effects of CO2 LASER. Ages ranged from 43 to 72 years. Nineteen patients were classified as Fitzpatrick III and 7 as Fitzpatrick II. Upper eyelid skin was examined before operation and after selective photothermolysis at 3 months, six months and one year. All post-operative evaluations showed consistent epidermal and dermal histological changes, such as the epidermal regeneration showing normal anatomy, mainly for evaluations around one year. Derma shows a dramatic transformation of neocollagen on medium and superficial derma, as well as the intense change (restructure) of elastic system and decrease of glucosamineglucans. The two laser passing were analyzed, as well as the whole healing process.
Keywords: Histological changes; eyelid skin; C02 1aser.
RESUMO
Para avaliar os efeitos histológicos da ação do LASER de CO2 sobre a pele palpebral, foram selecionados 26 pacientes, sendo 22 do sexo feminino e 4 do sexo masculino. A idade variou de 43 a 72 anos, sendo 19 deles classificados como Fitzpatrick III e 7 como Fitzpatrick II. A pele palpebral superior foi examinada antes da operação e após a fototermólise seletiva, aos três meses, seis meses e um ano. Todas as avaliações pós-operatórias mostraram consistentes modificações histológicas epidérmicas e dérmicas, como a regeneração epidérmica mostrando anatomia normalizada, principalmente nas avaliações em torno de um ano. A derme mostra dramática transformação de neocolágeno na derme superficial e média, como também intensa modificação (reestruturação) do sistema elástico e diminuição dos glicosaminoglicanos. As duas passagens do LASER foram analisadas, como também todo o processo de cicatrização.
Palavras-chave: Alterações histológicas; pele palpebral; laser de CO2.
ABSTRACT
The keloid is a pathology that affects exclusively the human being and can not be reproduced in any other animal species. Therefore, the knowledge regarding this pathology is limited and, several times, controversial. The ideal option for its treatment has not been completely established and there are several therapeutical methods. In this article we tried to develop a critical review of the literature making the current used methods evident and, therefore, discussing its major advantages and disadvantages.
Keywords: Keloid and treatment; keloid and infiltration; keloid and betatherapy; keloid and surgery; keloid and compression; scar management; laser
RESUMO
A cicatriz queloidiana é uma enfermidade que acomete apenas o ser humano, não sendo reproduzível em nenhuma outra espécie animal. Assim sendo, os conhecimentos a seu respeito são bastante limitados e muitas vezes controversos. A opção ideal para o seu tratamento ainda não foi completamente estabelecida, existindo inúmeros métodos terapêuticos. Neste artigo procuramos desenvolver uma revisão crítica da literatura, evidenciando os métodos utilizados atualmente e discutindo, assim, suas principais vantagens e desvantagens.
Palavras-chave: Quelóide e tratamento; quelóide e infiltração; quelóide e betaterapia; quelóide e cirurgia; quelóide e compressão; manejo de cicatrizes; laser
ABSTRACT
C02 laser Ultrapulse (Coherent Inc.) was used in 200 patients, from March 1996 to December 1997. In 180 patients (90%) full face laser resurfacing was done, and in 87 patients (48,3%) this procedure was associated with face lifting surgery. On the remaining 20 patients, CO2 laser was applied on other body areas, as dorsal hands, and for excision of warts and nevi. All patients submitted to laser resurfacing were previously treated during 1 to 2 months with retinoic acid and hydroquinone. The procedures were done under local anesthesia supervised by an anesthesiologist. A clear film dressing impregnated with silicone gel (Silon TSR, Bio-Med Sciences, Bethlehem, PA, USA) was used for 6 to 7 days and complete healing was observed in 7 to 10 days. Complications were exclusively dermatological, without relation to surgery. Ten consecutive patients have undergone skin biopsies for the study of the histological effects of the laser. Personal satisfaction, evaluated after 6 months to one year was excellent in 58%, good in 30%, regular in 10% and poor in 2%. In the last one, severe acne scars were observed, with indication to repeat the treatment.
Keywords: laser; resurfacing; facial rejuvenation.
RESUMO
Utilizamos o laser de CO2 Ultrapulse (Coherent Inc.,Palo Alto, CA) em 200 pacientes, de março de 96 a dezembro de 1997. O laser na face foi realizado em 180 pacientes (90%) sendo que em 87 pacientes (48,3%) esse procedimento foi associado a ritidoplastia. Nos 20 pacientes restantes, o laser de CO2 foi utilizado em outros segmentos do corpo, como no tratamento do dorso das mãos e na retirada de nervos e verrugas da face e de outras áreas do corpo. Todos os pacientes submetidos a laser de face foram preparados previamente por um período de 1 a 2 meses com ácido retinóico e hidroquinona. Os procedimentos foram realizados sob anestesia local e sedação, supervisionados por anestesiologista. Utilizou-se curativo semipermeável impregnado com silicone gel (Silon TSR) Bio-Med Ccinces, Bethlehem, PA, USA) por um período de 6 a 7 dias, com cicatrização completa após 7 a 10 dias. Todas as intercorrências pós-operatórias foram de origem dermatológica, sem qualquer complicação de ordem cirúrgica. Dez pacientes consecutivos foram submetidos a biópsia de pele para estudo histológico da ação do laser. A satisfação pessoal, avaliada após um período de 6 a 12 meses, foi ótima em 58%, boa em 30%, regular em 10% e insatisfatória em 2% dos casos. Os últimos foram seqüelas de acne grave com indicação de repetir o procedimento.
Palavras-chave: laser; resurfacing; rejuvenescimento facial
ABSTRACT
The authors report their experience with laserlipolysis, a surgical procedure that uses a Nd-YAG type laser with a 1064nm wavelength to treat localized adiposity, as an isolated or associated method to other traditional procedures. The physical principles of laser action on the adipocyte, anesthetic details, application technique, cases, advantaqes, limitations, disadvantages, complications and long-term results are analyzed. Moreover, the principle of action of laserlipolysis - photo-hyperthermia - on fat and adjacent tissue is demonstrated and analyzed by histological studies.
Keywords: Laserlipolysis; laser liposuction; Nd-Yag laser; peristaltic pump; photo-hyperthermia
RESUMO
Os autores relatam sua experiência com a laserlipólise, procedimento cirúrgico que utiliza um laser do tipo Nd-YAG com comprimento de onda de 1064nm no tratamento da adiposidade localizada, seja como método isolado ou associado a outros procedimentos tradicionais. São analisados os princípios físicos de ação do laser sobre o adipócito, detalhes anestésicos, a técnica de aplicação, casuística, vantagens, limitações, desvantagens, complicações e resultados a longo prazo. Além disso, é demonstrado e analisado, através de estudos histológicos, o princípio de ação da laserlipólise - foto-hipertermia - sobre a célula gordurosa e tecidos adjacentes.
Palavras-chave: Laserlipólise; lipoaspiração a laser; Nd-Yag laser; bomba peristáltica; foto-hipertermia
ABSTRACT
BACKGROUND: Reports on improvement in post-traumatic or pathological scars with the use of fractional CO2 laser (CO2F) conclude that it is a safe and effective technology, though used only in patients with phototypes II to III. The aim of this study was to evaluate the effectiveness of CO2F in patients with facial burn sequelae with phototypes III to VI. METHODS: A total of 14 patients (average age, 29 years) with facial burn sequelae and phototypes III to VI were subjected to a CO2F laser treatment. After 2 months, the burns were evaluated using a 6-parameter scale, including color, texture, hydration, surface irregularities, volume, and distensibility. RESULTS: The average durations of pain, edema, and hyperemia were 19 hours, 1.3 days, and 6.5 days, respectively. The fall of crusts was completed between 5 and 36 days with an average of 13.4 days. Two months after the session, 5 patients developed punctiform hypochromia in a checkerboard pattern corresponding to the points of laser ablation. The subjective satisfaction of the evaluators (i.e., both patients and physicians) with the treatment was 84.6%. The patients reported improvements in surface irregularities, distensibility, and skin texture (57% of the cases); hydration (43%); volume (28%); and color (14%). Meanwhile, the doctors reported improvements in surface irregularities and distensibility (43%). CONCLUSIONS: The treatment with CO2F laser with mild parameters was well tolerated and resulted in high satisfaction rates for patients with facial burn sequelae as well as improved skin texture, distensibility, and surface irregularities. The high incidence of hypopigmentation must be considered while prescribing CO2F laser treatment to patients with phototypes IV to VI.
Keywords: Burns/complications. Carbon dioxide. Laser therapy/methods.
RESUMO
INTRODUÇÃO: Relatos sobre melhora em cicatrizes pós-traumáticas ou patológicas com o uso do laser de CO2 fracionado (CO2F) concluem tratar-se de tecnologia segura e efetiva, apesar de utilizado apenas em pacientes com fototipos II a III. O objetivo deste estudo foi avaliar a eficácia do CO2F em pacientes com sequela de queimadura facial com fototipos III a VI. MÉTODO: No total, 14 pacientes, com média de idade de 29 anos, portadores de sequela de queimadura facial e fototipos III a VI, foram submetidos a uma aplicação do laser de CO2F. Após dois meses, a queimadura foi avaliada por meio de escala com 6 parâmetros: cor, textura, hidratação, irregularidades de superfície, volume e distensibilidade. RESULTADOS: A duração média da dor foi de 19 horas; do edema, 1,3 dia; e da hiperemia, 6,5 dias. A queda das crostas finalizou entre 5 dias e 36 dias, com média de 13,4 dias. Dois meses após a sessão, 5 pacientes evoluíram com hipocromia puntiforme no padrão quadriculado correspondente aos pontos de ablação do laser. A satisfação subjetiva dos avaliadores (pacientes e médicos) com o tratamento foi de 84,6%. Para os pacientes, houve melhora das irregularidades de superfície, da distensibilidade e da textura da pele (57% dos casos), da hidratação (43%), do volume (28%) e da cor (14%). Para os médicos, houve melhora das irregularidades de superfície e da distensibilidade (43%). CONCLUSÕES: O tratamento com laser de CO2F com parâmetros suaves foi bem tolerado e apresentou alto índice de satisfação em pacientes com sequela de queimadura facial, com melhora de textura, distensibilidade e irregularidades de superfície. A alta incidência de hipopigmentação é um fator a ser considerado na indicação do laser de CO2F em pacientes com fototipos IV a VI.
Palavras-chave: Queimaduras/complicações. Dióxido de carbono. Terapia a laser/métodos.
ABSTRACT
The exact indications of CO2 and Erbium laser are discussed with the objective of attaining better esthetic results and decreasing morbidity. Indications are based on the knowledge of tissue interaction of each laser; and technical details of the procedure based on 6-year experience in laser resurfacing.
Keywords: Laser; resurfacing; erbium versus CO2
RESUMO
As indicações precisas do laser dos tipos CO2 e Erbium são discutidas no intuito de apresentar um melhor resultado estético e diminuir a morbidade. As indicações são baseadas no conhecimento da interação tecidual de cada laser, apresentando detalhes técnicos de execução resultante de uma experiência de 6 anos em resurfacing laser.
Palavras-chave: Laser; resurfacing; erbium versus CO2
ABSTRACT
The authors describe the renew of the lower eyelid and a third half of the face through the mobilization of the eye's orbicularis oculi muscle of the lower eyelid on the blepharoplasty transconjunctival in 127 patients. It is discussed the technical evolution of the lower blefaroplasty moreover that it is currently possible to fulfill the necessary procedures through the transconjunctival approach that were before done only through transcutaneous, due to the introduction of the laser devices and of the new light and visualization mechanisms. It is described a new way of the miotomy and the elevation of the orbicular muscle of the eye, fulfilled maneuver essentially through transconjunctival and superior eyelid. The results are shown as satisfactory, and not occurring any definite complication.
Keywords: Blepharoplasty. Eyelids/surgery. Laser therapy. Oculomotor muscles/surgery.
RESUMO
Os autores descrevem a técnica de rejuvenescimento da pálpebra inferior e terço médio da face por meio da mobilização do músculo orbicular do olho da pálpebra inferior na blefaroplastia transconjuntival em 127 pacientes. É discutida a evolução técnica da blefaroplastia inferior, salientando que é atualmente possível realizar os procedimentos necessários pela via transconjuntival, que antes eram realizados somente pela via transcutânea, com auxílio de equipamentos de raios laser e de novos mecanismos de iluminação e visualização. São descritas uma nova abordagem da miotomia e a elevação do músculo orbicular do olho, manobras realizadas essencialmente pela via transconjuntival. São mostrados os resultados considerados satisfatórios, não ocorrendo nenhuma complicação definitiva.
Palavras-chave: Blefaroplastia. Pálpebras/cirurgia. Terapia a laser. Músculos oculomotores/cirurgia.
ABSTRACT
Introduction: Congenital vascular malformations up to know, have their best results with dye laser. We associate, in single equipment, long pulse Nd:Yag 1064 nm laser and intense pulsed light, for simultaneous treatment of the superficial and deep lesions. Methods: Seventeen patients were treated with the association described in an office basis with topical anesthesia only at monthly intervals by the same surgeon. The equipment utilized was the Synchro Platform (DEKA) with a 4 mm diameter spot size and fluency of 110 J/cm2 for the ND:Yag 1064 nm laser. For the intense pulsed light was utilized a 550 nm cutoff filter with a 46 x 10mm spot size and 21 J/cm2 as fluency, divided in two shots. Results were evaluated through clinical and photographical analysis together with the patient. Results: Among the seventeen patients treated, four results were considered excellent, nine, good, four, regular and none insufficient. Conclusion: The treatment can adequately be done in an office basis with topical anesthesia. Association of the two different technologies got a satisfactory result and it is probably one more technique for treating such lesions.
Keywords: Laser coagulation. Capillaries/Abnormalities. Hemangioma.
RESUMO
Introdução: As malformações vasculares congênitas do tipo capilar têm seus melhores resultados até o momento com o laser de corante. Nós associamos, em um único equipamento, o laser ND:Yag 1064 nm de pulso longo e a luz intensa pulsada, para o tratamento de lesões superficiais e profundas. Método: Dezessete pacientes foram tratados com a associação descrita em regime ambulatorial, com anestesia tópica, apenas em intervalos mensais, pelo mesmo cirurgião. O equipamento utilizado foi a plataforma Synchro (DEKA), com a ponteira de 4 mm de diâmetro e fluência de 110 J/cm2 para o laser ND:Yag 1064 nm. Para a luz intensa pulsada foi utilizado um filtro de corte de 550 nm, a ponteira de 46 x 10 mm e 21 J/cm2 de fluência, divididos em dois disparos. Os resultados foram avaliados clinicamente e por meio de fotografias, em conjunto com os pacientes. Resultados: Dentre os dezessete pacientes tratados, quatro resultados foram considerados excelentes, nove, bons, quatro, regulares e nenhum insuficiente. Conclusão: O tratamento pode ser realizado adequadamente com anestesia tópica em regime ambulatorial. A associação das duas tecnologias possibilitou a obtenção de resultados satisfatórios e, provavelmente, é mais uma técnica no tratamento destas lesões.
Palavras-chave: coagulação por laser. Capilares/anormalidades. Hermangioma.
ABSTRACT
The advent of Ultrapulse CO2 laser provided plastic surgeons with the opportunity to present their patients with the most simple and efficient treatment for skin lesions, including those from periocular region, periorbital and peribuccal wrinkles and frontal scars. The authors of this work present their experiences in 11 cases of resurfacing using CO2 laser. Material and methods used for skin preparation, type of anesthesia as well as a demonstration of how to do resurfacing and which intensity should be used in some cases are described; also, their experience with the exposed dressing) the postoperative and results are reported. Some photos of patients before and after laser application showing a satisfactory result for the physicians and a more pleasant life for the patients are shown on the conclusion.
Keywords: Resurfacing with C02 laser; laser in dark skins; facelift/full face; CO2 Tru-Pulse laser.
RESUMO
O advento do Laser Ultrapulsátil de CO2 deu aos cirurgiões plásticos a oportunidade de presentear seus pacientes com o mais simples e eficaz tratamento para lesões da pele, incluindo as da região periocular, as rugas periorbitais, peribucais e cicatrizes frontais. Os autores deste trabalho apresentam suas experiências em 111 casos de resurfacing com laser de CO2. São descritos os materiais e os métodos por eles utilizados para a preparação da pele, o tipo de anestesia, demonstrando-se como fazer o resurfacing e qual a intensidade que deverá ser utilizada em alguns casos; são relatados também suas experiências com o curativo expositivo, o pós-operatório e os resultados. Na conclusão, são exibidas algumas fotografias de pacientes antes e depois da aplicação do laser, demonstrando um resultado satisfatório para os doutores e uma vida mais agradável para os pacientes.
Palavras-chave: Resurfacing com laser de CO2; laser em peles morenas; facelift/full face; Tru-Pulse Laser de CO2.
ABSTRACT
Aesthetic plastic surgery has been greatly divulged in the last new years. Patient's expectations as to procedure outcomes are frequently unreal, showing lack of correct divulging of its limits and possibilities. This paper aims at discussing evaluation of results in aesthetic surgery, with the criteria for rhytidoplasty and mastoplasty as examples.
Keywords: Plastic surgery; aesthetic surgery; evaluation; face rejuvenescence; mastoplasty
RESUMO
A cirurgia plástica estética tem tido importante divulgação nos últimos anos. A expectativa dos pacientes quanto aos resultados dos procedimentos é freqüentemente irreal, demostrando a falta de correta divulgação de seus limites e possibilidades. O presente trabalho tem como finalidade discutir a avaliação dos resultados da cirurgia estética, exemplificando com os critérios utilizados para a ritidoplastia e mastoplastia.
Palavras-chave: Cirurgia plástica; cirurgia estética; avaliação; rejuvenescimento facial; mastoplastias
ABSTRACT
Although transconjunctval lower lid blepharoplasty was described many years long ago, it has only been in recent years, with the advent of laser; that its indication has become almost imperative. Thanks to this revival, transconjunctival blepharoplasty has been indicated in many situations. Authors have been using it more recently in most cases where resectioning or repositioning of the eyelid pads is necessary. Also, when indicated, scalpel or laser skin resection follows immediately. In this manner; since there are no lesions to nerve and muscle structures, mainly pre-tarsal structures, most of the complications of lower lid blepharoplasty can be avoided.
Keywords: Blepharoplasty; transconjunctival; laser
RESUMO
A blefaroplastia inferior transconjuntival foi descrita há muito tempo, mas somente nos últimos anos, com o advento do laser, é que sua indicação passou a ser quase imperativa. Devido a esse ressurgimento, a blefaroplastia transconjuntival voltou a ser indicada em muitas situações. Mais recentemente os autores a têm usado na maioria dos casos em que é necessária a ressecção de bolsas palpebrais ou seu reposicionamento. Quando também está indicada a ressecção de pele, esta é feita em seguida, com bisturi ou a laser. Assim, não há lesão das estruturas nervosas e musculares, principalmente pré-tarsais, o que evita a maioria das complicações das blefaroplastias inferiores.
Palavras-chave: Blefaroplastia; transconjuntival; laser
ABSTRACT
INTRODUCTION: On the basis of the available information, it is difficult to establish the appropriate dose of laser energy with which to obtain predictable results in laser-assisted liposuction. The purpose of this study is to evaluate the safety and efficacy of a 980-nm diode laser by using geometrical formulas. These formulas allow a precise quantification of tissue volume and, as a result, an estimation of the required laser dose. METHODS: This prospective study was conducted to evaluate 39 consecutive patients who underwent 980-nm diode laser-assisted liposuction (LAL) between November 2011 and November 2013. Four geometric patterns were used to determine the volume of fat before laser application. Ultrasonography was used to determine fat depth. Biopsies were conducted to determine immediate laser tissue reaction relative to laser dose. The aesthetic results and complications were recorded. Pearson's correlation coefficient was used to determine the relationship between the applied energy and tissue volume. RESULTS: 163 anatomic regions were treated with 980-nm LAL. The applied energy was relative to the volume in each area (R = 0.8786). 8 to 14 J/cm3 was the dose needed to produce effective hemostasis and minimal lipolysis. The results were very satisfactory in most cases. Biopsy results revealed lipolysis, ballooning, membrane rupture, and collagen band breakage; these changes varied with the radiation dose. The only complications were edema and transient bruising. CONCLUSION: The use of 3D geometrical models improved safety and efficacy of laser-assisted liposuction by permitting accurate determination of the tissue volume.
Keywords: Surgery, Plastic; Lipectomy; Mathematical models; Laser-therapy.
RESUMO
INTRODUÇÃO: Baseada na informação disponível encontram-se dificuldades em estabelecer a dose apropriada de energia a laser para alcançar resultados previsíveis na lipoaspiração a laser. Este estudo avaliou a segurança e eficácia do laser diodo 980-nm com uso de fórmulas geométricas. Essas fórmulas permitem uma qualificação precisa do volume do tecido, e como consequência, uma estimativa da dose de laser necessária. MÉTODOS: Tratase de estudo prospectivo que avaliou consecutivamente 39 pacientes submetidos a lipoaspiração a laser diodo (LLD) 980-nm entre Novembro de 2011 e Novembro de 2013. Quatro padrões geométricos foram adotados para determinar o volume de gordura antes da aplicação do laser. A ultrassonografia foi utilizada para determinar a profundidade da gordura. Foram realizadas biópsias para estabelecer a reação imediata do tecido ao laser relativo à dose do laser. Os resultados estéticos e as complicações foram documentadas. Utilizou-se o coeficiente de correlação de Pearson para determinar a relação entre a energia aplicada e o volume do tecido. RESULTADOS: Um total de 163 áreas anatômicas foram tratadas com LLD 980-nm. A energia aplicada foi baseada no volume de cada área (R = 0,8786). Foi necessário dose de 8 a 14 J/cm3 para produzir dose de hemostase efetiva, e lipectomia mínima. Os resultados foram muito satisfatórios na maioria dos casos. A biópsia relevou lipectomia, balonamento, ruptura de membrana, e rompimento de banda de colágeno; essas mudanças variaram com a dose de radiação. As complicações foram somente edema e hematoma transitório. CONCLUSÃO: O uso de modelos geométricos tridimensionais melhoraram a segurança e eficácia da lipoaspiração a laser, pois permitiram determinar o volume do tecido.
Palavras-chave: Cirurgia plástica; Lipectomia; Modelos matemáticos; Terapia a laser.
ABSTRACT
Introduction: Tissue repair and skin wound healing are complex processes that involve a series of dynamic events, including coagulation, inflammation, granulation tissue formation, wound contraction and tissue remodeling. The need to care for patients with difficulties in wound healing is a growing challenge and requires innovative strategies. One approach that stands out in the treatment of these injuries is low-power laser therapy. This study aims to demonstrate the applicability and results of a low-power laser therapy protocol in the management of skin wound healing.
Methods: This is a retrospective case review study reporting the use of laser in the treatment of skin wounds in a single center.
Results: We evaluated the results of the implementation of this low-power laser therapy protocol and reported five cases of patients with skin wounds that did not improve with conventional treatment. The lesions were located in different places, with lesions on the trunk, upper limb and lower limb. The five cases showed a good response to laser therapy, evolving with wound healing in a short period of time, with an average healing time of 48 days.
Conclusion: It is concluded that low power laser therapy when applied to skin wounds suggests a beneficial, promising action and has the potential to increase the therapeutic options available to the surgeon, however as we report a number of five cases, further studies are needed for check laser efficiency on wounds.
Keywords: Low-level light therapy; Wound healing; Cell proliferation; Laser therapy; Skin Diseases.
RESUMO
Introdução: O reparo tecidual e a cicatrização de feridas cutâneas são processos complexos que envolvem uma série de eventos dinâmicos, incluindo coagulação, inflamação, formação de tecido de granulação, contração da ferida e remodelação tecidual. A necessidade de cuidar de pacientes com dificuldades na cicatrização de feridas é um desafio crescente e requer estratégias inovadoras. Uma abordagem que se destaca no tratamento dessas lesões é a laserterapia de baixa potência. Este estudo tem como objetivo demonstrar a aplicabilidade e os resultados de um protocolo de laserterapia de baixa potência no manejo da cicatrização de feridas cutâneas.
Métodos: Trata-se de estudo retrospectivo de revisão de casos relatando o uso de laser no tratamento de feridas cutâneas em um centro único.
Resultados: Avaliamos os resultados do uso de laserterapia de baixa potência e relatamos cinco casos de pacientes com feridas cutânea que não apresentaram melhora com o tratamento convencional. As lesões eram localizadas em sítios diversos, com lesão no tronco, membro superior e membro inferior. Os cinco casos apresentaram uma boa resposta com a terapia a laser, evoluindo com a cicatrização das feridas em um curto período de tempo, com período médio de cicatrização de 48 dias.
Conclusão: Conclui-se que a laserterapia de baixa potência quando aplicada sobre feridas cutâneas sugere uma ação benéfica, promissora e tem um potencial para aumentar as opções terapêuticas disponíveis para o cirurgião, porém, como relatamos um número de cinco casos, há necessidade de mais estudos para verificar a eficiência do laser em feridas.
Palavras-chave: Terapia com luz de baixa intensidade; Cicatrização; Proliferação de células; Terapia a laser; Pele.
ABSTRACT
BACKGROUND: Laserlipolisis is a surgical procedure that uses a Nd:YAG laser for treatment of localized lipodistrophies as an isolated or associated method. The present study aims to evaluate the results of laserlipolisis with Nd:YAG laser based on a retrospective survey. METHOD: Retrospective evaluation with 766 patients in 792 procedures from August 2001 to August 2011. RESULTS: Mean follow-up was 6.5 years (78 months). The most frequent indication was lipodistrophy of abdomen, flancs and dorsum. A satisfaction questionnaire showed 96% of patients satisfied. CONCLUSIONS: Laserlipolisis brought several benefits to obtained results as better accomodation of skin, skin retraction, less edema and bleeding, favoring treatment of more extensive areas as well as more complex regions as gynecomastias, secondary liposuction and dorsal regions, with better postoperative recovery.
Keywords: Nd:Yag laser. Laserlipolisis. Laser-assisted liposuction. Nd:Yag laser.
RESUMO
INTRODUÇÃO: A técnica de lipoaspiração a laser (laserlipólise) é um procedimento cirúrgico que se utiliza de um aparelho de laser do tipo Nd:YAG (neodymium-doped:yttrium-aluminum garnet) para o tratamento de adiposidades localizadas, seja como método isolado ou associado a outros procedimentos tradicionais de cirurgia plástica. Este estudo teve por objetivo avaliar os resultados do uso da lipoaspiração com laser Nd:YAG no tratamento das lipodistrofias, baseado em avaliação retrospectiva de 10 anos. MÉTODO: Estudo retrospectivo, que avaliou 766 pacientes (792 cirurgias) submetidos ao procedimento entre agosto de 2001 e agosto de 2011. RESULTADOS: O seguimento médio dos pacientes foi de 6,5 anos (78 meses). A indicação mais frequente para o procedimento foi para o tratamento das regiões de abdome, flancos e dorso. O resultado do questionário de satisfação demonstrou que 96% dos pacientes consideraram o resultado satisfatório CONCLUSÕES: Aaplicação da laserlipólise trouxe vários benefícios adicionais aos resultados, como a constatação de melhor acomodação e retração da pele, menor edema e sangramento, possibilidade de tratamento de áreas de maior complexidade para a lipoaspiração convencional, como ginecomastia, dorso e lipoaspiração secundária, além de melhor recuperação pós-operatória.
Palavras-chave: Laserlipólise. Lipoaspiração a laser. Laser Nd:Yag.
ABSTRACT
Objective: To present the changes in the size of the skin, according to security parameters for the stored energy in laser lipolysis procedures in several anatomical regions, using diode laser device, with dual wavelength. Method: A prospective, cross-sectional case series included 41 patients undergoing laser lipolysis of deformities of the body contour, between June 2008 and July 2010. The laser lipolysis was performed with diode laser device, accessory to the Aspire platform, brand Slim-Lipo® (Palomar Medical Technologies, Inc), with two independently controlled wavelength, one operating at 924 nm and another at 975nm, using tips in three sizes, according suitability for small, medium or large areas of adipose tissue treated. Measurements were compared the Wilcoxon test at a significance level of 0.05. Results: All patients had significant retraction of the skin independent of the surgical site (face, breast adiposity, arm or waist), in the absence of complications. Conclusion: The laser diode proved capable to promote progressive and persistent skin retraction, complying with the proposed security parameter.
Keywords: Lipectomy/ methods. Laser Therapy. Surgery, Plastic.
RESUMO
Objetivo: Apresentar as alterações na dimensão da pele, obedecendo aos parâmetros de segurança relativos à energia acumulada em procedimentos de laserlipólise em diversas regiões anatômicas, empregando aparelho de laser diodo com duplo comprimento de onda (924 e 975 nm). Método: Um estudo prospectivo, transversal, tipo série de casos, incluiu 41 pacientes submetidos à laserlipólise para tratamento de deformidades do contorno corporal, entre junho de 2008 e julho de 2010. A laserlipólise foi realizada com aparelho de laser diodo com dois comprimentos de ondas independentemente controlados, um operando a 924 nm e outro a 975nm, empregando ponteiras de três tamanhos, segundo adequação a pequenas, médias ou grandes áreas de tecido adiposo tratado. As aferições foram comparadas pelo teste de Wilcoxon, em nível de significância de 0,05. Resultados: Em todos os pacientes, houve retração significante da pele, independente do sítio cirúrgico (face, adiposidade mamária, braços ou cintura), em ausência de complicações. Conclusão: O laser diodo mostrou-se capaz de promover retração da pele, progressiva e persistente, obedecido o parâmetro de segurança proposto.
Palavras-chave: Lipectomia/ métodos. Terapia a Laser. Cirurgia Plástica.
ABSTRACT
Introduction: The benefits of the interruption of the tabagism previously the surgery not yet well are defined. Objective: The objective of this study was to assess the influence of the interruption of exposure to nicotine on a cutaneous flap, by laser flowmetry and picrosirius. Method: 56 rats were used, divided in groups and administred nicotine 2mg/kg/day, 0,5mg/kg/day and just saline solution. Animals were subdivided according to exposure time to the drug, that is, in one and two weeks' interruption prior to surgery, and constant treatment up to ten days after surgical procedure. They were organized in seven groups. A cutaneous flap was performed on the animals' dorsum and on the tenth day of PO the laser flow measurement and the formation of collagen and its differentiation in mature and immature by picrosirius. Results: The flowmetry was bigger in the control and the groups of low doses. The interruption of the drug only showed increase of the flowmetry in the groups of low dose. The ratio of mature colageno for immature was inversely proportional to the dose and the time of exposition to the drug. Conclusion: The nicotine harms the evolution of a remnant cutaneous of the back of rats, being dependent of the dose. The interruption time modifies the flowmetry and the ratio of formation of colageno, when given in low dose.
Keywords: Nicotine. Laser-doppler flowmetry. Surgical flaps. Picrates.
RESUMO
Introdução: Os benefícios da interrupção do tabagismo previamente à cirurgia ainda não estão bem definidos. Objetivo: Avaliar a influência da interrupção da nicotina em um retalho cutâneo, por meio da fluxometria a laser e com picrosirius. Método: Foram utilizados 56 ratos, divididos em grupos expostos à nicotina 2mg/kg/dia, e a 0,5mg/kg/dia, e somente à solução salina. Os animais foram subdivididos de acordo com o tempo de exposição, em 1 e 2 semanas de interrupção previamente à cirurgia e tratamento ininterrupto até 10 dias do pós-operatório. Após o 10º dia, foram avaliados os retalhos cutâneos do dorso do animal, por meio de fluxometria a laser, formação de colágeno, e diferenciação em maduro e imaturo, pelo picrosirius. Resultados: A fluxometria foi maior no controle e nos grupos de baixas doses. A interrupção da droga mostrou aumento da fluxometria somente nos grupos de baixa dose. A proporção de colágeno maduro para imaturo foi inversamente proporcional à dose e ao tempo de exposição à droga. Conclusões: A nicotina prejudica a evolução de um retalho cutâneo do dorso de ratos, sendo dependente da dose. O tempo de interrupção altera a fluxometria e a proporção de formação de colágeno, quando dada em baixa dose.
Palavras-chave: Nicotina. Fluxometria por laser-doppler. Retalhos cirúrgicos. Picratos.
ABSTRACT
Stretch marks are scars on the dermis that cause patients to be self-conscious and that pose challenges in clinical treatment. They result from rapid stretching of the skin and often observed in adolescents, pregnant women, obese individuals, and people experiencing rapid change in weight. They initially appear as erythematous and edematous plaques (red striae), and as they mature, they become whitish and atrophic (striae alba) due to the degradation and reorganization of elastin and collagen fibers. Currently, laser treatment is an emerging noninvasive therapeutic modality that is successful in reducing the excessive vascularization of red striae and in stimulating the production of collagen and elastin in the alba. In the present literature review, PubMed was searched for articles on the treatment of atrophic striations with laser that were published from January 2000 to December 2016. The authors have found 28 articles that met the inclusion criteria. Only few randomized clinical trials have evaluated the long-term efficacy and safety of energybased treatments. Based on clinical cases and systematic reviews, both ablative and non-ablative fractional lasers caused moderate improvement in the appearance of distending striae. The 1540-nm non-ablative fractional laser is more likely to be considered the first-line treatment for stretch marks. However, future studies should focus on research with longer duration, studies with objective and standardized measures for the evaluation of results, such as biopsy results, and molecular studies showing an increase in elastic and collagen fibers that correlate to the improvement in the clinical appearance of the striae after using lasers with systematized parameters.
Keywords: Stretch marks; Laser therapy; Dermatology; Skin; Reconstructive surgical procedures
RESUMO
Estrias de distensão são cicatrizes na derme que constrangem os pacientes e oferecem desafios ao tratamento clínico. Resultam do rápido estiramento da pele, frequentemente presente em adolescentes, gestantes, obesos e indivíduos com rápida mudança de peso. Apresentam-se inicialmente como placas eritematosas e edemaciadas (estrias rubras) e, conforme amadurecem, tornam-se esbranquiçadas e atróficas (estrias albas), devido à degradação e reorganização de fibras de elastina e de colágeno. Atualmente, lasers representam modalidade terapêutica não invasiva emergente, que já demonstrou sucesso na redução da vascularização excessiva das estrias rubras, e no estímulo à produção de colágeno e elastina nas albas. Foi realizada revisão da literatura no PubMed referente ao tratamento de estrias atróficas com laser de janeiro de 2000 até dezembro de 2016. Os autores encontraram 28 artigos que se enquadraram nos critérios de inclusão. Existem poucos estudos clínicos randomizados avaliando a eficácia ao longo prazo e a segurança de aparelhos à base de energia. Baseado em casos clínicos e revisões sistemáticas, ambos os lasers - ablativo e não ablativo - fracionados demonstraram melhora modesta do aspecto das estrias distensas. Há tendência a sugestão do laser não ablativo fracionado de 1.540 nanômetro como sendo opção de primeira escolha para tratamento das estrias distensas. Porém, estudos futuros devem focar em desenhos de pesquisa com duração mais longa, medidas objetivas e padronizadas da avaliação dos resultados como biópsias e estudos moleculares, que demonstrem aumento nas fibras elásticas e fibras de colágeno, que correlacionem com a melhora do aspecto clínico das estrias após aplicação de lasers com parâmetros sistematizados.
Palavras-chave: Estrias de distensão; Terapia a laser; Dermatologia; Pele; Procedimentos cirúrgicos reconstrutivos
ABSTRACT
The use of synthetic implants as dermal facial fillers may cause adverse reactions, which may occur owing to inappropriate techniques or the intrinsic nature of the product. Polymethylmetacrylate (PMMA) microspheres is one of the materials used. This is a report of a case of nodular reaction in the face of a female patient, after 15 years of Artecoll® implant use, a product composed of microspheres of bovine collagen and PMMA. She was treated with a carbon dioxide laser. The delayed effect caused by the application of synthetic implants in the face, such as Artecoll®; the limiting factors of conventional resection with a scalpel and scissors; and the option for the use of carbon dioxide laser are outlined. Highlighted are the ablative and precise functions in the removal of compromised tissue, the preservation of healthy and noble tissues, the high hemostatic potential, and low morbidity, with smaller scars, edema, ecchymosis, and rapid postoperative recovery.
Keywords: Laser; Adverse reaction; Polymethylmetacrylate.
RESUMO
O uso de implantes sintéticos para o preenchimento na face pode acarretar reações adversas. Estas podem ocorrer da má técnica ou decorrentes do produto. Entre os materiais utilizados, está o polimetilmetacrilato (PMMA) na forma de microesferas. Trata-se do relato de um caso de reação nodular na face de paciente do sexo feminino, após 15 anos do uso de implante de Artecoll®, produto composto por microesferas de colágeno bovino e PMMA. Foi tratada com laser de gás carbônico. Evidenciou-se o efeito tardio provocado da aplicação de implantes sintéticos na face como o Artecoll®, os fatores limitantes para ressecção convencional com bisturi e tesoura, e a opção da utilização do laser de CO2. Destacando-se a função ablativa e precisa na remoção do tecido comprometido, a preservação dos tecidos sadios e nobres, o grande potencial hemostático e a baixa morbidade, com menor cicatriz, edema, equimose e a rápida recuperação pós-operatória.
Palavras-chave: Laser; Reação adversa; Polimetilmetacrilato.
All scientific articles published at www.rbcp.org.br are licensed under a Creative Commons license