INTRODUCTION
Liposuction is one of the main procedures performed by plastic surgeons. Over time,
a number of important innovations have been made to traditional liposuction, which
have resulted in greater comfort for the surgeon as well as better results1.
Historically, several approaches have been used to remove adipose tissue during liposuction2. Over time the procedure has been refined, and improvements were made to improve
various aspects of the procedure, such as the surgical technique, the cannulas employed
and the use of adjuvant devices.
In a sense, this will be a continuous journey. There are no particular objectives,
but only goals. Traditional liposuction remains a somewhat strenuous procedure with
limited tactical variations. Therefore, initiatives capable of generating a reduction
in the load and mechanical stress or amplifying the handling of the subcutaneous cell
tissue variations can help optimize lipsuction3.
The use of ultrasound in surgical procedures is not new. Zocchi, in 19954, became a pioneer in applying ultrasound to emulsify fat selectively so as to contour
the body. The first-generation devices for ultrasonic-assisted liposuction were developed
by SMEI, Italy. The SMEI appliance consisted of solid 4-6 mm probes that emulsified
the fat at a frequency of 20 kHz.
Second-generation devices were introduced in the 1990s, such as the Lysonix 2000 (Lysonix
Inc. Carpenteria, CA). In this case, emulsification and aspiration occurred simultaneously
through a "golf-tee" and "bullet design" cannula at a frequency of 22.5 kHz. During
the same period, the Mentor Corporation introduced its body contouring devices, Mentor
Contour Genesis Devices, with a hollow 3.0 mm and 5.1 mm cannula at a frequency of
27 kHz. The excessive transfer of energy to the tissues combined with the elimination
of the protective layer from the simultaneous extraction resulted in significant complications5.
The popularity of ultrasonic-assisted liposuction decreased by the end of the 1990s.
In 2001, Sound Surgical Technologies introduced VASERTM (Vibration Amplification of Sound Energy at Resonance), a third-generation device
that was created to improve safety by reducing the energy transferred to the tissues
while maintaining efficacy6.
VASERTM has a solid probe that emulsifies fat efficiently at 36 kHz while preserving nearby
tissue. The tunable nature of the system allows almost all areas of the body to be
treated safely and effectively. Currently, VASERTM is considered the gold standard in high-definition body contour7.
In VASERTM, the resonance uses a frequency of ٣٦ kHz, which is close to the resonance of fat.
For this reason, a lower energy is transferred to other tissue. Moreover, adipose
cells are much larger than other adjacent tissues, such as blood vessels, nerves,
and connective tissue and, therefore, are more susceptible to ultrasonic energy8.
The appliance uses 2.2-4.5 mm diameter probes with grooves near the tip to increase
the energy transmission efficiency and fragmentation of fat. A larger probe results
in more ultrasonic energy being dispersed. The device also has a pulsed ultrasonic
energy delivery mode. This method uses high vibration frequencies with non-continuous
activation, decreasing the total tissue energy applied while maintaining efficacy9.
OBJECTIVE
This study aimed to evaluate the safety of a third-generation ultrasonic device, VASERTM, with liposuction surgeries to improve body contour. The data obtained in this study
were compared with the existing literature.
METHODS
The study consisted of a retrospective review of medical records of patients who underwent
a liposuction procedure with the aid of VASERTM, from January 2015 to June 2017, at the Santa Monica Hospital Center in Erechim,
Rio Grande do Sul, Brazil.
The data and the methods used were approved by the Ethics Committee of the Santa Monica
Hospital, Erechim/RS, Brazil. Record 002/2018.
Selection of patients as well as inclusion and exclusion criteria
The patients included in the study were male or female, aged over 18 years, with localized
excess subcutaneous fat. The exclusion criteria were:
Patients with a limiting clinical condition;
Women in the first year after birth, were pregnant or lactating;
A body image disorder;
Inflammatory conditions of the skin in the target area of surgical therapy;
Obesity (BMI > 30);
Concomitant abdominoplasty surgery.
We included 76 patients in our study. Females were the most common gender, with 74
patients, accounting for 97.36% of the sample. Two male patients (2.67%) underwent
liposuction with VASERTM. The average age of the women in the study was 39 years old (21 to 65 years) and
37 years old for men. The mean BMI of the sample was 24.64 kg/m2 (Table 1).
Table 1 - Procedures performed.
Procedure performed |
Number of Patients |
Liposuction in the abdominal region, flanks, and dorsum. |
31 |
Liposuction in the abdominal region and dorsum with the placement of breast implants. |
20 |
Liposuction in the abdominal region, flanks, and dorsum with mastopexy. |
8 |
Liposuction in the abdominal region and mastopexy with the placement of implants. |
5 |
Liposuction in the abdominal region, flanks, and dorsum with mastopexy and placement
of implants.
|
3 |
Liposuction in the abdominal region with the placement of breast implants. |
3 |
Liposuction in the lateral face of the thigh with the placement of breast implants. |
2 |
Liposuction in the abdominal region. |
1 |
Liposuction of the abdominal region, flanks, and dorsum with correction of gynecomastia. |
2 |
Liposuction of the lateral face of the thighs. |
1 |
Table 1 - Procedures performed.
Evaluation Criteria
Given that liposuction was performed, we used the routine criteria mentioned in the
current literature regarding surgical complications. These can be classified according
to the occurrence period.
Preoperative evaluation
All patients were evaluated before surgery by the anesthesiology team of the Santa
Monica Hospital Center. Laboratory and complementary exams were requested in accordance
with their age and comorbidities. Patients were under general anesthesia for the procedure
(Table 2).
Table 2 - Complications.
Perioperative complications (0-48 h)
|
Postoperative complications in recent (1-7 days) |
Complications in the late postoperative period (1 week-3 months) |
Skin necrosis |
Cellulite |
Seroma |
Injury due to the cannula, portal or endpoint |
Paresthesia, transient or permanent alteration in sensitivity |
Prolonged Edema |
Anesthetic Complications |
Hyperpigmentation |
Fibrosis |
Hypopigmentation |
|
VASERTM Technology
Ultrasonic technology is based on the conversion of electrical energy in sound vibrating
waves through a handpiece. This vibration moves through a titanium probe at a constant
frequency of 36 kHz. The interaction of probe designs with tissue sound reverberation
regulates the effectiveness of the system2.
Subcutaneous tissue, which is in a humid environment, receives the sound energy coming
from this probe. There are two operating mechanisms. The first, and less frequent,
breaks down the cell membrane by direct impact. The second is called cavitation. The
vibratory frequency of the sound wave produces compression and rarefaction forces
(cavitation), which are issued by its distal rings, with small air bubbles forming.
These bubbles will gradually increase in diameter and simultaneously englobe the adipocytes
in their interior until they rupture. The energy released by the rupture of the microbubbles
releases the adipocytes from the tissue microarchitecture8.
Cavitation and the mechanical rupture of tissue are selective, i.e., the diameter
of fat cells are larger than the adjacent tissue microarchitecture (blood vessels,
muscle fibers, connective tissues). Therefore, these remain intact8.
The appliance has enough power and accuracy to treat different body areas without
the need to use loading to overcome areas of tissue resistance6.
The VASERTM is composed of an integrated system, formed by a display (ultrasound), handpiece,
probes (special atraumatic rods), aspiration tower using the Ventx system (ventilated
suction system), integrated irrigation system, and drive pedals. This entire apparatus
brings together all the necessary equipment to perform liposuction surgery in an organized
manner. This helps to ensure safety, comfort, and practicality in using the device8 (Figure 1).
Figure 1 - VASERTM System (Vibration Amplification of Sound Energy at Resonance). Source: The Author (2018)
Figure 1 - VASERTM System (Vibration Amplification of Sound Energy at Resonance). Source: The Author (2018)
Skin Ports
Protective parts used in surgical accesses are routinely inserted into the gluteus
sulcus, anterior and posterior axillary topographies, pubic region, navel, and inframammary
sulcus. They prevent and protect these areas from thermal lesions occurring due to
ultrasonic vibrations and trauma of repetitive movements (Figure 2).
Figure 2 - Portals with a suitable design for varying the diameters of the ultrasonic probe.
Source: The Author (2018)
Figure 2 - Portals with a suitable design for varying the diameters of the ultrasonic probe.
Source: The Author (2018)
Handpiece
Sound pulse conduction instrument connected directly to the ultrasound with inserts
for the probes. They require continuous revision of "wrench" fitting to dissipate
the sound wave vertically (Figure 3).
Figure 3 - Handpiece with probes and a wrench. Source: The Author (2018)
Figure 3 - Handpiece with probes and a wrench. Source: The Author (2018)
Cannulas
Atraumatic cannula system for collecting emulsified tissues, with amplitudes and configurations
for variable anatomical units6 (Figure 4).
Figure 4 - Set of VentxTM cannulae, from top to bottom: Adaptation handle, 3.0 mm infiltration
cannula, 3.0 mm liposuction cannulae, 3.7 mm, 4.6 mm short, 4.6 mm long, 5.0 mm basket
cannula, 4.6 mm curved cannula, baby armpit cannula, 3.0 mm Toledo cannula, 3.0 mm
lower limbs cannula and 4.0 mm cannula for fat grafting. Source: The Author (2018)
Figure 4 - Set of VentxTM cannulae, from top to bottom: Adaptation handle, 3.0 mm infiltration
cannula, 3.0 mm liposuction cannulae, 3.7 mm, 4.6 mm short, 4.6 mm long, 5.0 mm basket
cannula, 4.6 mm curved cannula, baby armpit cannula, 3.0 mm Toledo cannula, 3.0 mm
lower limbs cannula and 4.0 mm cannula for fat grafting. Source: The Author (2018)
Probes
The equipment has probes of different diameters with grooves near the tip to transmit
power. The larger the diameter of the probe, the greater the dissipated energy. Formatting
for all anatomical units is available7 (Figure 5).
Figure 5 - VASERTM ultrasonic probes. From left to right: 4.5 mm probe, 3.7 mm probe/2 rings,
3.7 mm probe/3 rings, arrow probe and 2.9 mm probe/3 rings and Saturn probe. Source:
The Author (2018)
Figure 5 - VASERTM ultrasonic probes. From left to right: 4.5 mm probe, 3.7 mm probe/2 rings,
3.7 mm probe/3 rings, arrow probe and 2.9 mm probe/3 rings and Saturn probe. Source:
The Author (2018)
Presurgical marking
During the physical assessment, an analysis of the distribution of adipose tissue
in different parts of the body is established. This includes documenting the thickness
of the abdominal and trunk fat, as well as muscle mass.
We use routine VASERTM instead of a body contouring liposuction procedure as the surgical protocol indications
are identical, with no exceptions.
We begin with the patient in the orthostatic position, and the first anatomical landmarks
delimited are the areas in which the withdrawal of 100% of the lamellar layer of subcutaneous
tissue is planned. Then, the rectus abdominis muscles, the anterosuperior iliac crests,
and the inguinal ligaments are identified. The topography of the gluteus maximus muscles,
the sacral concavity, and the transition between the upper and lower back following
the lumbosacral fascia syncope are also delimited (Figure 6).
Figure 6 - Blue: delimitation of deep extraction. Black: delimitation of strategic anatomical
points. Lilac: transitional areas of superficial relief. Green: Region of total extraction
of lamellar layer and partial extraction of areolar compartment. Red: alert area to
avoid excessive resection. Source: The Author (2018)
Figure 6 - Blue: delimitation of deep extraction. Black: delimitation of strategic anatomical
points. Lilac: transitional areas of superficial relief. Green: Region of total extraction
of lamellar layer and partial extraction of areolar compartment. Red: alert area to
avoid excessive resection. Source: The Author (2018)
Surgical technique
The preparation of the patient in the surgical center begins with anesthesia; we prefer
using general anesthesia. After that, a delayed bladder catheter, socks, and intermittent
venous compression equipment for the lower limbs and body warming systems are prepared.
The surgery is standardized in three stages: infiltration, emulsification, and aspiration.
Positioning
The patient is positioned in ventral decubitus with ample exposure of anatomical units
and, after the end of this step, changed to dorsal decubitus. In both positions, strategic
portals are placed in masking areas (back: intergluteal sulcus and in posterior axillary
fold topography. Abdomen: in the pubic region, umbilical scar, inframammary topography,
and anterior axillary topography). These accesses receive, after infiltration, protectors
(skin ports) to slide the probes, avoiding adjacent thermal injuries (Figure 7).
Figure 7 - Patient in dorsal decubitus with safety portals in the pubic area, the upper edge
of navel and inframammary sulcus (prorings). Source: The Author (2018)
Figure 7 - Patient in dorsal decubitus with safety portals in the pubic area, the upper edge
of navel and inframammary sulcus (prorings). Source: The Author (2018)
Infiltration
The super humid infiltration technique is used for both superficial and deep regions.
The routine volume used is 1:1 (infiltrate volume/aspirate volume), with a warm saline
solution and epinephrine (1 ampoule for each 1000 mL saline). The solution is inserted
through the previously made portals. The adjacent region needs to be kept moist during
the surgical stages with a physiological solution to dissipate better the thermal
energy generated by the device and friction.
Emulsification
The introduction of the probes follows the working movement practiced in traditional
liposuction, that is, back and forth movements, without force, feeling the device
cross the tissues. This process begins with the lamellar layer and ends in the areolar
compartment.
The recommended duration of VASERTM, according to guidelines from the manufacturer, is approximately 1 minute for each
100 mL of infiltrated solution, to generate the feeling of a "loss of resistance".
Areas of higher fat content can be addressed with a larger diameter probe and with
the appliance power up to 80% in continuous mode. In less thick areas, such as the
waist, smaller probes, 2.9 mm, with 3 rings (greater lateral dispersion of energy)
and the appliance power up to 60%, in pulsed mode, is recommended.
The superficial use of the VASERTM is performed with a 2.9 mm probe, 3 rings, and in pulsed mode. This step is responsible
for the retraction of the skin, offsetting the areas of myofascial flaccidity, especially
the hypogastrium. The duration of the emulsification stage in our routine is approximately
30-45 minutes in each decubitus (Table 3).
Table 3 - VASERTM system probes.
Density |
Volume |
Probe |
Model |
Frequency |
Smooth |
Medium/Large |
3.7 mm (3 grooves) |
Continuous |
70-80 |
Smooth |
Small |
2.9 mm (3 grooves) or 3.7 mm (2 grooves) |
Pulsed or continuous |
70-80 |
Slightly fibrous |
Medium/Large |
3.7 mm (2 grooves) |
Continuous |
80-90 |
Slightly fibrous |
Small |
2.9 mm (3 grooves) or 3.7 mm (1 grooves) |
Continuous |
80-90 |
Very fibrous |
Medium/Large |
3.7 mm (2 grooves) or 3.7 mm (1 groove) |
Continuous |
80-90 |
Very fibrous |
Small |
2.9 mm (3 grooves) |
Continuous |
80-90 |
Table 3 - VASERTM system probes.
LIPOSUCTION
Lamellar layer
The aspiration of fat follows the conventional technique of tissue collection, that
is, "fan-shaped" and with a "spread hand," feeling the movement of the cannula. Tissue
resistance is minimal, given that the density of the tissue is markedly decreased
by emulsification. The cannulae used have a diameter of 3.0 mm, 3.7 mm, and 4.0 mm.
The aspiration of the tissue is always initiated by the deep layer, with greater diameter
cannulae (3.7 mm and 4.0 mm).
Areolar layer
The superficial emulsification and liposuction are performed selectively on the edges
of the muscle groups (alba line and inguinal ligament). Thinner cannulas are essential.
Transition
This phase is used to ease the transition between the muscle groups and the sharp
edges in the superficial liposuction, promoting the withdrawal of excess muscle fat,
with an improvement of the definition of the anatomical landmarks superficially. The
goal is the demarcation of the lateral edges of the muscle groups, with the maintenance
of a pinch test of around 1 cm.
Superficial subdermal liposuction
This consists of aspirating superficial subdermal fat through thin 3.0 mm and 2.0
mm cannulae. By reducing the fat just below the skin, it is possible to obtain effective
skin retraction. The Ventx system enables continuous negative pressure in the liposuction
cannula. Consequently, the aggression of the cannula when sliding through tissue,
in particular to the subdermal plexus, does not damage it (Figure 8).
Figure 8 - Example of liposuction with approaches of the various levels of lamellar fat anatomical
relief, joint thinning of the subcutaneous cellular tissue in areas of tendon encounters
and the adjacent anatomical transitions
Figure 8 - Example of liposuction with approaches of the various levels of lamellar fat anatomical
relief, joint thinning of the subcutaneous cellular tissue in areas of tendon encounters
and the adjacent anatomical transitions
After the end of the liposuction step, we introduce a continuous aspiration drain
(PortoVac) in the lumbosacral and suprapubic regions.
Postoperative management
Patients use a compressive modeling mesh and venous return pump throughout their hospital
stay. On the second postoperative day, patients begin daily postoperative lymphatic
drainage for at least 10 days.
RESULTS
In this retrospective study, we performed an analysis of the medical records of patients
that underwent liposuction using the VASERTM system to assess the surgical complications found. We then compared our findings
with the existing literature.
Our main goal was to evaluate the safety of the VASERTM device. Other data such as the volume infiltrated and aspirated solution, the mode
of energy used (pulsed or continuous), and application time were also collected. Some
results will be mentioned in general terms.
The maximum volume did not exceed 4,000 mL in any patient. In all cases, the supernatant
fat exceeded 80% of the total of the aspirate volume (Figure 9). In some cases, this index was above 93%.
Figure 9 - Emulsified fat. In our sample, the supernatant fat exceeded 80% of the total lipoaspirated
volume. Source: The Author (2018) .
Figure 9 - Emulsified fat. In our sample, the supernatant fat exceeded 80% of the total lipoaspirated
volume. Source: The Author (2018) .
Among the patients analyzed in this study, 71 (93.42%) presented no complications
related to postsurgical liposuction with the aid of the VASERTM. Five cases (6.57%) presented complications (Table 4).
Tabela 4 - Complications.
Complications |
Number of patients |
% |
Prolonged edema |
1 |
1.31 |
Thermal injury in insertion portal |
1 |
1.31 |
Hyperpigmentation |
2 |
2.67 |
Epidermal lesion |
1 |
1.31 |
Cellulite |
0 |
0.00 |
Anesthetic complications |
0 |
0.00 |
Fibrosis |
0 |
0.00 |
Changes in sensitivity |
0 |
0.00 |
Seroma |
0 |
0.00 |
Total |
5 |
6.57 |
Tabela 4 - Complications.
The hyperpigmentation and thermal injury events in the insertion portal occurred in
the same patient. The remainder of the complications presented in different patients.
The transmission of ultrasonic energy to the tissue can cause a lesion at the insertion
portal site or terminal damage, according to the excessive static maintenance of the
probes9, or a wear to the skin port, as in our sample. Excision and suturing were performed.
The epidermal lesion also occurred in one flank after use of the VASERTM. The patient presented progressive improvement with intensive care of the affected
area. Dyschromia and hypo- or hyperpigmentation of the skin from using ultrasonic
devices has been reported in the literature9,10. Hyperpigmentation may occur due to the release of hemosiderin and its deposit, causing
alteration in color, or the use of modulators that improperly compress tissues, possibly
associated with the pathophysiology of this complication11. The two patients affected recovered after compression relief.
There was no formation of seroma in our study. Only one case of prolonged edema (1.31%)
occurred, which was resolved with intensive conservative therapy. We believe that
the low rates found in our sample are due to the standardization of lymphatic drains
and the systematic use of vacuum drains (Figure 10).
Figure 10 - A and C. Preoperative photographs of a 38-year-old male with gynecomastia. B and D. Postoperative
photographs 8 weeks after surgical treatment with VASERTM assisted ultrasonic liposuction.
Infiltration of 300 mL of saline solution per side, with a 2.9 mm probe for 3 minutes
and 7 seconds on the right and 3 minutes and 20 seconds on the left. 70% power was
used in pulsed mode. Total volume aspirated of 520 mL (90% of emulsified fat tissue)
Figure 10 - A and C. Preoperative photographs of a 38-year-old male with gynecomastia. B and D. Postoperative
photographs 8 weeks after surgical treatment with VASERTM assisted ultrasonic liposuction.
Infiltration of 300 mL of saline solution per side, with a 2.9 mm probe for 3 minutes
and 7 seconds on the right and 3 minutes and 20 seconds on the left. 70% power was
used in pulsed mode. Total volume aspirated of 520 mL (90% of emulsified fat tissue)
DISCUSSION
In this study, 76 patients were submitted to ultrasonic liposuction assisted with
VASERTM for the treatment of lipodystrophies. The results indicate that the emulsification
of subcutaneous tissue through the continuous and pulsed ultrasound mode was efficient
and safe. The aspirated material contained more than 80% fat supernatant, reaching
93% in some cases. Blood losses were not relevant, considering that the characteristics
of infranatant liquids aspirated were insignificant and, by association, no clinical
management by depletion was performed. These findings are in agreement with the comparative
study conducted by Garcia Junior and Nathan, in ٢٠٠٨12, which concluded that the use of third-generation ultrasonic devices leads to 20%
less blood loss than other techniques12.
Jewell, Fodor, Pinto, and Al Shammari, in 20028, reviewed the literature with a statistical analysis of the surgical complications
related to liposuction. Ninety-three articles were selected, of which 14 were statistically
evaluated. An average of 13.5% of unwanted events was observed in the studied literature.
In our sample, we obtained 6.57% of postoperative complications, thus assuring the
safety of this technology.
In 19989, Rohrich et al. sought to evaluate the experience of using an ultrasonic liposuction
device in with 114 consecutive patients. In their sample, five complications were
observed: one case of dysesthesia, which the author believed to have been due to an
excessively prolonged use of the device, three cases of abdominal seroma (with need
for aspiration and compression) and one case of thermal lesions in the insertion portal
of the cannula, likely due to an improper driving technique. The author emphasizes
the need for maintenance of a damp environment and constant movement of the cannula.
These are positive results when compared to traditional liposuction. The author relates
the complications with the learning curve.
Dixit and Wagh, in 201311, conducted a literary review on postoperative problems in liposuction, reporting
a rate of 18.7% cutaneous hyperpigmentation in manipulated areas. They state that
the leading cause of this is the inadequate approach of the areolar layer with consequent
lesion of the subdermal plexus. In our study, we found 2.67% events of this nature.
The low incidence of this complication is significant because, in at least 92.1% of
the cases, the areolar layer was affected in one anatomical unit with fat emulsification
and subsequent liposuction. The small amount of adverse events in superficial liposuction
is the most important finding in terms of safety10. Since the technical foundations underlying conventional liposuction is the deep manipulation
of fat tissue in a humid environment, a thickness of approximately 1 cm of tissue
is left to prevent contour irregularities and/or superficial devitalization3.
Nagy and Vanek Junior, in 201213, performed a multicenter randomized study comparing traditional liposuction and the
ultrasonic method on contralateral anatomical regions of the same patients, with female
patients aged between 20 and 48 years old.
Increased tissue retraction occurred in 53% per cc aspirated when the VASERTM was used, which was statistically significant and clinically relevant14.
Based on these assumptions of higher skin retraction and manipulation of the areolar
layer without tissue devitalization, it is possible to obtain greater muscle definition
in patients with favorable physical characteristics6.
We agree with this premise despite having subjective contextualization. When performing
techniques that aim for greater definition of compartments, ensuring safety using
appropriate probes, selecting the correct pulsed or intermittent energy dispersion,
and providing super humid superficial conditions, deep infiltration, and lamellar
aspiration with subsequent areolar collection are essential6 (Figure 11)..
Figure 11 - A. Photographs in the immediate postoperative period of a 32-year-old woman, after
VASERTM assisted liposuction in the abdominal region, flanks, and dorsum with a deep
approach to the entire circumference and superficial and deep manipulation in the
muscle intersection areas. B, C, and D. Pinch test with different thicknesses according to the surgical plan
Figure 11 - A. Photographs in the immediate postoperative period of a 32-year-old woman, after
VASERTM assisted liposuction in the abdominal region, flanks, and dorsum with a deep
approach to the entire circumference and superficial and deep manipulation in the
muscle intersection areas. B, C, and D. Pinch test with different thicknesses according to the surgical plan
Hoyos and Millard, in 20072, stated that VASERTM technology enables surgeons to perform superficial lipoplasty techniques efficaciously.
The first- and second-generation ultrasonic devices were associated with severe burns
and necroses when they dispersed their energy on the surface. These issues are not
prevalent in the third-generation technology15,16,17,18, which can be seen in our study.
CONCLUSION
The existing literature, as well as our analysis, shows that VASERTM liposuction procedures can improve body contour safely, with low complication rates.
The potential complications from ultrasonic devices, such as tissue ischemia or necrosis,
can be mostly attributed to the device being used inappropriately. It is of fundamental
importance to standardize care concerning the insertion portals and to ensure an adequate
amount of infiltrating solution is used as well as that the device is applied for
the appropriate amount of time and areas.
Therefore VASERTM is a potential tool that is safe for surgeons to use in body contouring procedures.
*Felipe Massignan and the other authors did not receive any financial compensation
or benefits from the company responsible for the device mentioned in this article.
COLLABORATIONS
FM
|
Analysis and/or data interpretation, Conception and design study, Conceptualization,
Data Curation, Formal Analysis, Investigation, Methodology, Project Administration,
Realization of operations and/or trials, Resources, Supervision, Validation, Visualization,
Writing - Original Draft Preparation, Writing - Review & Editing
|
REFERENCES
1. Fodor PB, Cimino WW, Watson JP, Tahernia A. Suction-assisted lipoplasty: physics,
optimization, and clinical verification. Aesthetic Surg J. 2005 May;25(3):234-246.
2. Hoyos AE, Millard JA. VASER-assisted high-definition liposculpture. Aesthetic Surg
J. 2007 Nov/Dec;27(6):594-604.
3. Markman B, Barton Junior FE. Anatomy of the subcutaneous tissue of the trunk and lower
extremity. Plast Reconstr Surg. 1987 Aug;80(2):248-254.
4. Zocchi ML. Ultrasound-assisted lipoplasty. Adv Plast Reconstr Surg. 1995;11:197-221.
5. Scuderi N, Paolini G, Grippaudo FR, Tenna S. Comparative evaluation of traditional,
ultrasonic, and pneumatic assisted lipoplasty: analysis of local and systemic effects,
efficacy, and costs of the methods. Aesthetic Plast Surg. 2000 Nov/Dec;24(6):395-400.
6. Hoyos AE. High definition liposculpture. Presented in the XIII International Course
of Plastic Surgery. Bucaramanga, Colombia; 2003.
7. Cimino WW. Ultrasonic surgery: power quantification and efficiency optimization. Aesthetic
Surg J. 2001 May;21(3):233-41.
8. Jewell ML, Fodor PB, Pinto EBS, Al Shammari MA. Clinical application of VASER-assisted
lipoplasty: a pilot clinical study. Aesthetic Surg J. 2002 Mar;22(2):131-46.
9. Rohrich RJ, Beran SJ, Kenkel JM, Adams Junior WP Jr, DiSpaltro F. Extending the role
of liposuction in body contouring with ultrasound-assisted liposuction. Plast Reconstr
Surg. 1998 Apr;101(4):1090-102;discussion:1117-9.
10. Rocha RP, Rocha ELP, Pinto EBS, Sementilli A, Nakanishi CP. Cutis marmorata resemblance
after liposuction. Aesthetic Plast Surg. 2005 Jul/Aug;29(4):310-2.
11. Dixit VV, Wagh MS. Unfavorable outcomes of liposuction and their management. Indian
J Plastic Surg. 2013 May/Aug;46(2):377-92.
12. Garcia Junior O, Nathan N. Comparative analysis of blood loss in suction-assisted
lipoplasty and third-generation internal ultrasound-assisted lipoplasty. Aesthetic
Surg J. 2008 Jul/Aug;28(4):430-5.
13. Nagy MW, Vanek Junior PF. A multicenter, prospective, randomized, single-blind, controlled
clinical trial comparing VASER-assisted Lipoplasty and suction-assisted Lipoplasty.
Plast Reconstr Surg. 2012 Apr;129(4):681-9e.
14. Rocha RP, Pinto EBS, Cardoso RA, Nardino EP, Rodrigues L, Tonetti RL. Morphological
aspects of dorsum and sacrum skin as related to liposuction. Aesthetic Surg J. 2000
May/Jun;24(3):171-8.
15. Zucowski ML, Ash K. Ultrasound-assisted lipoplasty learning curve. Aesthetic Surg
J. 1998 Mar;18(2):104-10.
16. Troillius C. Ultrasound-assisted lipoplasty: is it really safe?. Aesthetic Plast Surg.
1999;23:307-311.
17. Scheflan M, Tazi H. Ultrasonically assisted body contouring. Aesthetic Surg J. 1996
Jun;16(2):117-122.
18. Grolleau JL, Rougé D, Chavoin JP, Costagliola M. Severe cutaneous necrosis after ultrasound
lipolysis. Medicolegal aspects and review. Ann Chir Plast Esthet. 1997;42(1):31-36.
1. Centro Hospitalar Santa Mônica, Centro Hospitalar, Erechim, RS, Brazil.
Corresponding author: Felipe Massignan Rua Itália, 878, Sala 105, Espírito Santo, Erechim, RS, Brazil. Zip code: 99711-010.
E-mail: felipemassignan@me.com
Article received: April 4, 2019.
Article accepted: October 20, 2019.
Conflicts of interest: none.